What to pay attention to when starting up dental equipment after the New Year holidays?

Dental equipment needs to be activated after the holidays in order to serve the clinic better.
I think most dentists will be a little maladjusted after a long holiday of New Year. But did you know? Dental equipment also needs an “adaptation process” after a long vacation.
How to do avoid damaging dental equipment when stating up after a long vacation?Don’t worry, we will show you solutions all one by one. In order to ensure that the dental equipment can work smoothly after the holidays, the following tips are warmly advised.I hope they are useful for all of you.
When turning on dental imaging equipment
- Check the stability of the voltage of the use of the environment and the state of the line, do not turn on the dental equipment together with the downtime for many days to ensure that the line is good. The voltage of the imaging equipment should be stable at 220V±5%.
- The correct order of power on: open the UPS or voltage regulator → open the computer and other control systems → open the dental device host → wait for the dental device self-test.
- Disinfection and wiping of long-lived dental equipment parts (such as bite jaw lever, etc.) before power on and use to ensure a healthy and hygienic environment.
- Check the emergency switch of the dental equipment after turning on the machine, should not be pressed to ensure that it has been popped; ensure that the connection line between the control system and the dental equipment is normal; ensure that the ambient temperature is appropriate, the temperature is controlled at 10-40℃, relative humidity Keep at 35%-75%.
When turning on dental chair
We know that the dental chair is used very frequently, and this type of dental equipment is a combination of electricity and water, so please pay special attention to the following issues we mentioned.
- Check the voltage and line safety first before starting the dental unit, keep the voltage stable at 220V±11V.
- Check the water line and water pressure, whether the air pressure of air compressor reaches the standard value; the water pressure is 200Kpa-400Kpa. the air pressure should not be lower than 500kpa.
- Before turning on the dental unit, drain the water for 2 minutes to avoid the dirty water settling in the pipe.
- After turning on the dental unit, flush the inside of the hose of apparatus like dental handpieces and the pilot system for 2 minutes first.
Something important
- Rinse the inside of the suction hoses and the suction system by pumping one cup of water per hose. Purpose: To ensure that the dental chair remains in optimal working condition and provides a strong guarantee for therapeutic use.
- Do manual deep sterilization: inside of dental instrument hoses, drinking water system, suction system, outside of suction hoses, spittoon, etc. Purpose: Reasonable periodic maintenance can ensure that the dental chair is as good as new day after day and delay the service life.
“Holiday Syndrome”
The above methods are good solutions to the “Holiday Syndrome” of dental equipment, what about us? In order to eliminate the “Holiday Syndrome”, we can go to bed early, wake up early, make positive self-adjustment, seriously put into work, and start our journey to the new year.
If you have any questions, you can always talk to us, or if you need advice on how to import dental equipment from China, you can always find us, read our previous articles, you can also check out the relevant dental equipments like dental low-speed handpiece,intra-oral camera,apex loactor,endodontic sonic irigator activator and so on.