Why Does The Drill of Implant 20:1 Contra Angle Get Stuck?

In this article, we will analyze the problem why the driller of implant 20:1 contra angle get stuck.
Recently when a customer purchased China dental handpiece from us,he asked Why does The Drill of Implant 20:1 Contra Angle Get Stuck?
Below we analyze the problem that why the driller of Implant 20:1 Contra Angle Get Stuck.
First, we take a normal implant low speed handpiece, which can easily hold the drill without push button pressure and we can hear the sound of the drill being held.

Then we took another implant 20:1 contra angle that could not hold the drill, but the drill could be removed without push button pressure.

Then we took off the back cover of the implant 20:1 contra angle which could not hold the drill. Try to see if the dental drill can be stuck in the absence of the back cover, under normal circumstances can be stuck.


So we can find that the reason why the dental drill can not be stuck is that the implant 20:1 contra angle may have been dropped to the ground, the back cover was deformed after the collision, long-term in a state of being pressed open, so the dental drill has been concave part of the top, thus causing the inability to clamp the dental drill.

We often sell Implant 20:1 Contra Angle, and we also do the dental handpiece repair business for dental high speed handpiece, dental low speed handpiece and implant contra angle.If you have dental handpieces need to be repair, you can buy dental handpiece spare parts and repair tools from us.